Reacondicionado o Usado Monitor Datascope Passport

Reacondicionado o Usado Monitor Datascope Passport

Web ID: 100
Condición: Retired

0 Artículo está en su carrito Ver carrito ahora

Términos y Condiciones del Listado

All product and company names are trademarks of their respective holders. Use of them does not imply any affiliation with or endorsement or sponsorship by them.

Details

El Monitor Datascope Passport mide ECG, Respiración, NIBP (Presión Arterial no Invasiva), SPO2, y Temperatura. El Passport tiene una pantalla electro-luminiscente grande para vista fácil de cada ángulo. DRE vende el Passport remanufacturado.

Especificaciones:

  • Dimensiones: 9.5" x 13" x 8"
  • Peso: 13 libras máximas sin baterías
  • Consumo de Energía: 15 vatios típicos
  • Entrada de Voltaje: 100, 120, 220, 240 VAC ± 10%
  • Frecuencia de Energía: 47 a 65 Hz

ECG

  • Derivaciones: I, II, III
  • Sensibilidad de Pantalla: 0.25, 0.5, 1.0, 2.0, 3.0, 4.0 cm/mV ± 10%

Contador de Ritmo Cardiaco

  • Rango: 30-300 bpm adulto/niño. 30-350 bpm neonatal
  • Precisión: 30-250 bpm ± 3bpm o 3%
  • Actualización de Pantalla: 2 segundos ± 0.1 seg

NIBP (Presión Arterial No Invasiva)

  • Técnica: osilométrico
  • Rango sistólico: Modo adulto/pediátrico: 40-250 mmHg, Modo neonatal: 15-150 mmHg.
  • Rango diastólico: Modo adulto/pediátrico: 30-200 mmHg, modo neonatal: 10-150 mmHg.
  • Precisión sistólica: error media menos de 7 mmHg, desviación estándar menos de 8 mmHg (adulto/pediátrico). Error media menos de 3 mmHg, desviación estándar menos de 13 mmHg (neonatal).
  • Precisión diastólica: Error media menos de 7 mmHg, desviación estándar menos de 8 mmHg (adulto/pediátrico). Error media menos de 4 mmHg, desviación estándar menos de 7 mmHg (neonatal).
  • Rango de contador de ritmo cardiaco: 30-250 bpm.
  • Precisión de contador de ritmo cardiaco: ±3 bpm o ±3%

Oximetro de Pulso

  • Precision: 70-100%, ±2%; 60-69%, ±4%; 0-59%, no especificada
  • Tiempo de Responder: 13 segundos

Grabador (opcional)

  • Velocidad: 50, 25, 12.5, 6.25, 3.125 mm/seg ±5%
  • Atención: 6.25 y 3.125 son para la respiración solamente
  • Tamaño de Tabla: 50 mm ±2mm

Requisitos de Batería

  • 2 baterías
  • Voltaje: 12 VDC
  • Capacidad: 2.3 AH cada una
  • Duración de Operación: 2 horas con baterías completamente cargadas
  • Tiempo de Cargar: 16 horas máximo
  • Aviso de batería baja: 11.6 ± 3.5%
  • Limite de batería baja: 11.0 ± 2.8%

Specifications

General

  • Height: 9.5 inches
  • Width: 13 inches
  • Depth: 8 inches
  • Weight: 13 lbs. maximum without batteries
  • Power Consumption: 15 watts typical
  • Voltage input: 100, 120, 220, 240 VAC ± 10%
  • Mains frequency: 47 to 65 Hz

ECG

  • Leads: I, II, III
  • Display sensitivity: 0.25, 0.5, 1.0, 2.0, 3.0, 4.0 cm/mV ± 10%
  • Frequency response to screen: ESU filter out: 0.5 to 40 Hz, -3db max at 1 mVpp input; ESU filter in: 0.5 to 20 Hz -3db max at 1 mVpp input

Heart Rate Meter

  • Range: 30-300 bpm adult/child. 30-350 bpm neonatal
  • Accuracy: 30-250 bpm ± 3bpm or 3%, whichever is greater.
  • Pacer rejection: rejects all pulse of amplitude ± 2.0 mV to ± 700 mV and duration 0.1 ms to 2 ms with no tail. Rejects all pulses ± 2.0 mV to ± 700 mV and duration 0.1 ms to 2 ms with 100 ms T.C. tail <0.8 mV or 25 ms T.C. tail <0.4 mV.
  • Display update: 2 seconds ± 0.1 sec

NIBP

  • Technique: oscillometric
  • Systolic range: adult/ped mode: 40-250 mmHg, neonate mode: 15-150 mmHg.
  • Diastolic range: adult/ped mode: 30-200 mmHg, neonate mode: 10-150 mmHg.
  • Systolic accuracy: mean error less than 7 mmHg, standard deviation less than 8 mmHg (adult/pediatric). Mean error less than 3 mmHg, standard deviation less than 13 mmHg (neonatal).
  • Diastolic accuracy: mean error less than 7 mmHg, standard deviation less than 8 mmHg (adult/pediatric). Mean error less than 4 mmHg, standard deviation less than 7 mmHg (neonatal).
  • Heart rate meter range: 30-250 bpm.
  • Heart rate meter accuracy: ±3 bpm or ±3%, whichever is greater.

Pulse Oximetry

  • Accuracy (standard deviation): 70-100%, ±2%; 60-69%, ±4%; 0-59%, unspecified.
  • Response time: 13 seconds to 95% of final step change % SpO2 value from 60-95% at 72 bpm

Recorder (optional)

  • Speed: 50, 25, 12.5, 6.25, 3.125 mm/sec ±5%
  • Note: 6.25 and 3.125 are for respiration only
  • Chart size: 50 mm ±2mm

Battery Requirements

  • 2 sealed lead batteries
  • Voltage: 12 VDC
  • Capacity: 2.3 AH each
  • Run time: 2 hours fully charged battery
  • Recharge time: 16 hours maximum
  • Low battery warning: 11.6 ± 3.5%
  • Low battery cut off: 11.0 ± 2.8%