The MAC 5500 HD Resting ECG System is a premier ECG system, delivering advanced disease management capabilities through one of the industry-leading collections of algorithms and advanced networking.
The MAC 5500 HD represents a new definition of ECG speed and accuracy, and offers:
Display
Display Type - 10.4 in (264 mm) diagonal graphics backlit color AM LCD
Display Resolution - 640 x 480 pixels with waveform enhancement
Display Data - Heart rate, patient name, ID, clock, waveforms, lead labels, speed, gain and filter settings, warning messages, prompts, and help messages
Writer
Writer Technology - Thermal dot array
Writer Speeds - 5, 12.5, 25, and 50 mm/s
Number of Traces - 3, 6, 12, or 15 user selectable
Writer Sensitivity/Gain - 2.5, 5, 10, 20, 10/5 (split calibration) mm/mV
Writer Speed Accuracy - ±2% Writer
Amplitude Accuracy - ±5%
Writer Resolution - Horizontal 1000 dpi @ 25 mm/s, 200 dpi vertical
Paper Type - Thermal, Z-fold, perforated, fan fold, 300 sheets/pack
Paper Size A Size - 8.5 in x 11 in, (214.6 mm x 280 mm) A4 Size: 8.27 in x 11.7 in (210 mm x 297.5 mm)
Keyboard
Type - Sealed elastomer with soft function keys, alphanumeric keys, writer controls, and TrimPad cursor controls
Electrical
Power Supply - AC or battery operation
Voltage - 100 to 240 VAC
Current - 0.5A @ 115 VAC, 0.3A @ 240 VAC, typical, 0.85A max
Frequency - 50 to 60 Hz
Battery Type - User replaceable, 18V @ 3.5 AH ±10% rechargeable NiMH
Battery Capacity - 100 single page reports, (typical) or six hours continuous display (without printing)
Battery Charge Time - Approximately 4.5 hours from total discharge (with display off)
Vectorcardiography
Report Formats - Vector loops of component vectors (P, QRS, ST-T)
Sensitivity - 20, 40, 80, or 160 mm/mV
Time Resolution - 2 ms
Hi-Res Late Potential Analysis and P-Wave Signal– Averaged ECG
Sensitivities:
Raw Data Template - 20 mm/mV
Average Beat - 20 mm/mV and 50 mm/mV
Filtered Signals and Vector Magnitude - 1 mm/µV
Analysis Sampling Rate - 1,000 samples/second/channel
Digital Sampling Rate - 16,000 samples/second/channel
High/Low Pass Filters Special filter using Fast Fourier Transform (FFT)
Physical Specifications
Height - 3.7 in (9.4 cm)† with display closed
Width - 15 in (38.1 cm)†
Depth - 13.8 in (35.1 cm)†
Weight - Approximately 6.8 kg (15 lbs)† including battery, without paper
Environmental Specifications
Temperature
Operating - 50°to 104° F (10° to 40° C)
Transport/Storage - –40°to 158° F (–40° to 70° C)
Humidity
Operating - 20% to 95% RH non-condensing
Transport/Storage - 15% to 95% RH non-condensing
Pressure
Operating - 700 to 1060 hPA
Transport/Storage - 500 to 1060 hPA
Magnetic Card Reader Specifications
Character Set ANSI/ISO ALPHA alphanumeric characters and ANSI/ISO BCD (subset of ASCII [ISO 646 IRV:1991])
Bar Code Scanner Specifications
Symbologies Code 39 (extended), PDF-417, Code 128, Data Matrix, Interleaved 2 of 5
Modular MAC Trolley Dimensions
Height - 37 in (94 cm)
Width - 19 in (47 cm)
Depth - 27 in (69 cm)
Height with Acquisition module holder - 59 in (150 cm)
Weight - 66 lbs. (30 kg)††