The Datascope Passport XG is a standard in portable and bedside vital signs monitoring. the Passport XG's incredible 52 square-inch display is available in your choice of an Active Matrix Color TFT screen or a bright, high-contrast monochrome electroluminescent screen. Larger NIBP numbers of Automatic Waveform Display make reading your patient's data easier.

Features

  • 3/5 Lead ECG
  • Printer
  • Non-invasive blood pressure
  • Pulse oximetry
  • Battery back up for transport

Specifications

Display

21.1 × 15.8 cm

ECG

  • Leads I, II, III, aVR, aVL, aVF and V.
  • Defibrillator overload protection: Withstand Lown and trapezoidal or Edmark waveform up to 400 joules applied to 100 ohms load and electrode at 20 second intervals.
  • Recovery time: time for trace to return to screen from defibrillator moo pulse is 3 seconds maximum.

IBP

  • Range to digital display; systolic/diastolic mean -20 to +300 mmHg.
  • Accuracy: 2 mmHg or 2%, whichever is greater to numerical display position.
  • Zero range: ±120 mmHg minimum at nominal gain.

Heart Rate

  • Range - 30-300 bpm for adult/child, 30-350 bpm for neonatal.
  • Accuracy - 30-250 ±3 bpm or±3%, whichever is greater.

NIBP

  • Technique: oscillometric.
  • Systolic range: 50 to 250 mmHg for adult/ped mode, 15 to 150 mmHg for neonatal mode.
  • Diastolic range: 30 to 200 mmHg for adult/ped mode, 10 to 150 mmHg for neonatal mode.
  • Systolic accurancy: mean error less than 7 mmHg, standard deviation less than 8 mmHg (adult/ped), mean error less than 4 mmHg, standard deviation less than 13 mmHg (neonatal).
  • Diastolic accuracy: mean error less than 7 mmHg, standard deviation less than 8 mmHg (adult/ped), mean error less than 4 mmHg, standard deviation less than 7 mmHg (neonatal).
  • Heart rate meter range: 30-250 bpm.
  • Heart rate meter accuracy: ±3 bpm or ±3%, whichever is greater.

Temperature

  • Scale: selectable °C or°F.
  • Range: 15°C to 45°C, 59°F to 113°F.
  • Accuracy: ±0.1°C to 45°C or ±0.2°F 59 to 113°F

Respiration (ECG)

  • Range: 4-200 pm.
  • Accuracy: ±2% or ± bpm, whichever is greater

Recorder

Speed - 50, 25, 12., 6.25, 3.125 mm/sec ±5%

CO2

  • Range: 0-14.8 vol%, 0-14.8 kPa, 0-11 mmHg, 0-111 Torr.
  • Rise Time: <400ms.
  • Accuracy: ±(0.2% vol% to 2% of reading)

O2

  • Range: 0-100 vol%.
  • Rise time: <400 ms.
  • Accuracy: ±(0.2% vol% + 2% of reading)

N2O

  • Range: 0-100% N2O.
  • Rise time: <450 ms.
  • Accuracy: ±(2 vol%+ 2% of reading)

Battery

  • Type: 2 sealed lead acid batteries.
  • Voltage: 12 VDC nominal (@2.3A load).
  • Capacity: 2.3 AH each.
  • Run time: 1.5 hours from full charge of new battery at 25°C with SpO2, 1 NIBP every 15 minutes, no recording.
  • Recharge time: 16 hours.
  • Low battery warning: 11.6V, ±3.5%.
  • Low battery cut off: 11.0V ±2.8%