Marca: Equipo Médico GE
Web ID: 977
Condición: Refurbished

0 Artículo está en su carrito Ver carrito ahora

Términos y Condiciones del Listado

All product and company names are trademarks of their respective holders. Use of them does not imply any affiliation with or endorsement or sponsorship by them.

Details

Sólo el DASH 2000 Pro proporciona a los médicos con el rendimiento estándar de oro GE Medical Systems Information Technologies a un precio tan asequible. Esta combinación única de rendimiento y valor hace que el DASH 2000 Pro sea ideal para salas de procedimientos especiales y consultorios médicos.

Características

Estándar de Oro en Precisión NIBP

  • único monitor con todas las funciones y tecnología DINAMAP® incorporada.
  • Sistema produce consistentemente las determinaciones de NIBP más precisas y confiables disponibles en un monitor de cabecera.
  • La deflación escalada con tecnología de punta patentada ayuda a manejar el artefacto.
  • Ideal para NICU, PICU y adultos hipertensos.

Estándar de Oro en detección de arritmias

  • Detección de arritmia multi-lead simultáneas EK-Pro® establece el estándar para la sensibilidad y especificidad en un monitor de paciente.
  • Múltiples conductores aseguran un monitoreo ininterrumpido y ayudan a detectar eventos localizados que de otro modo podrían perderse.
  • Actualización incremental ayuda a eliminar el ruido, para un seguimiento preciso de los cambios sutiles, progresistas en formas de ritmo.
  • Correlación continua proporciona la detección QRS precisa en presencia de ruido o algún artefacto.
  • El análisis contextual ayuda a reducir las tasas de falsas alarmas y mejorar el reconocimiento de la arritmia en relación con el ritmo único de un paciente dado.

Fácil de Usar

  • Comparte una interfaz de usuario común con todos los monitores GE Medical Systems Information Technologies para la rápida y confiable transición de un monitor al siguiente.
  • Estructura de menús intuitivos simplifica la operación y minimiza el entrenamiento.
  • Conectores comunes con monitores SOLAR® y Tram® permiten la transición sin problemas por si acaso cambia la condición del paciente para permitir la monitorización más precisa.

Specifications

Physical Display

  • Height: 10.25" (26 cm)
  • Depth: 8" (20 cm)
  • Width: 11" (28 cm)
  • Weight: 12 lbs (5.5 kg) maximum
  • Size: 5.8" (16 cm) diagonal
  • Type: monochrome LCD; optional color LCD
  • Resolution: 320 x 240 pixels
  • Number of traces: 3
  • Number of seconds/trace: 3.8 at 25 mm/sec
  • Sweep speed: 25 mm/sec (with erase bar)
  • Information window: display non-real-time information without obstructing the display of real-time information
  • Display organization: prioritized by parameter

ECG

  • Leads: I, II, III, V, aVR, aVL, and a VF.
  • Lead fail: identifies failed lead.
  • Alarms: user-selectable upper and lower limits.
  • Heart rate range: 30 to 300 bpm

Respiration

  • Measurement technique: impedance variation detection.
  • Respiration rate range: 0 to 200 breathes per minute.
  • Alarms: user-selectable upper and lower respiration rate limits and apnea limit.

IBP

  • Number of channels: 1.
  • Transducer sites: arterial, pulmonary arterial venous, left arterial, intracranial and special.
  • Range: -25 mmHg to 300mmHg.
  • Alarms: user-selectable upper and lower limits for systolic, diastolic and mean pressures

Temperature

  • Number of channels: 1.
  • Range: 0°C to 45°C (32°F to 113°F).
  • Resolution: ±0.1°C (0.18°F).
  • Accuracy: ±0.1°C (0.18°F).
  • Alarms: user-selectable upper and lower limits

NIBP

  • Measurement technique: oscillometric.
  • Displayed parameters: systolic, diastolic and mean pressures, pulse rate, time of last measurement.
  • Measurement modes: manual, auto and start in adult ICU and OR modes; a/manual and auto in neonatal mode.
  • Heart rate detection: 30 to 300 beats per minute.
  • Total cycle time: 20 to 40 seconds typical.
  • Automatic cycle times: 0 to 24 hours.

Pulse Oximetry

  • Parameters monitored: arterial oxygen saturation (SpO2) and peripheral pulse rate (PPR).
  • SpO2 range: 50 to 100%.
  • PPR range: 30 to 300 beats per minute (±3 beats per minute).
  • SpO2 accuracy: ±2% (70 to 100% SpO2),±3% (50 to 69% SpO2).
  • PPR accuracy:±3 beats per minute

Alarms

  • Categories: patient status and system status.
  • Priority: 4 levels - crisis, warning, advisory, message.
  • Notification: audible and visual.
  • Setting: default and individual.
  • Silencing: 1 minute, current alarm only.
  • Pause: 5 minutes in adult ICU mode, 3 minutes in neonatal ICU mode, and 5, 15 minutes or permanent pause in OR mode.
  • Volume: default 70%, 70 dB measured at 1 meter.

Controls

  • Trim Knob® control.
  • Five hard keys: silence alarm, graph go/stop, NBP go/stop, function and power on/off.

Electrical

  • Power requirements: 90 to 264 VAC 50/60 Hz 500 mA.
  • Power consumption: 40 Watts (fully loaded).
  • Cooling: convection. Heat dissipation: 135Btu/br maximum