Details

Features multiple temperature control modes, each controlled by micro-computer

Features

  • Use AC power and DC power alternatively, and DC 12V or DC 24V power in an ambulance
  • Set temperature, air temperature, baby temperature and internal battery voltage can be displayed separately
  • Can be equipped with >37°C temperature set function
  • Failure alarm indication
  • Double wall hood with side door for pulling out the bassinet
  • Natural air flow humidity
  • Adjustable height

Configurations available with:

  • Main body, including the hood, controller, bassinet, internal batter and observation lamp
  • Oxygen cylinders
  • Oxygen supply system
  • Skin temperature sensor
  • IV pole
  • Mattress
  • Adjustable stand
  • >37° C override temperature mode (optional)

 

Specifications

General

  • AC power supply: AC110V - 240V/50-60Hz
  • DC power supply: DC12V/10A or DC24V/6A
  • Input power: 400VA
  • Air-temperature control range: 25° - 37°C
    • - Override mode: 37° - 38°C
  • Baby temperature control range: 34° - 37°C
    • - Override mode: 37° - 37.5°C
  • Skin temperature sensor accuracy: ≤0.3° C
  • Temperature variability: ≤1.0° C
  • Temperature uniformity: ≤1.5° C (mattress is on the horizontal position)
  • Noise inside hood: ≤55dB(A)
    • - Environment noise ≤45dB(A)
  • Failure alarm: Over-temperature alarm, temperature deviation alarm, sensor alarm, motor fan failed alarm, power failure alarm
  • Operating at temperature equilibrium: 90 min (Set temp: 36° C | Ambient temperature: 15° C)

Environmental Temperature

  • Operating range: 10° C ~ 30° C
  • Set temperature should be 3° C higher than the environment temperature
  • Ambient air movement rate: <1.0 m/s

Package