Datascope Passport EL Multi-Parameter Patient Monitor

Datascope Passport EL Multi-Parameter Patient Monitor

Web ID: 100
Condition: Retired

0 item is in your cart. View cart now

Listing Terms & Conditions

All product and company names are trademarks of their respective holders. Use of them does not imply any affiliation with or endorsement or sponsorship by them.

Details

The Datascope Passport monitor measures ECG, Respiration, NIBP, SPO2, and Temperature. The Passport has a large electroluminescent screen for easy viewing from any angle. This monitor is operated with easy to use, single function keys. Avante sells the Passport refurbished.

Specifications

General

  • Height: 9.5 inches
  • Width: 13 inches
  • Depth: 8 inches
  • Weight: 13 lbs. maximum without batteries
  • Power Consumption: 15 watts typical
  • Voltage input: 100, 120, 220, 240 VAC ± 10%
  • Mains frequency: 47 to 65 Hz

ECG

  • Leads: I, II, III
  • Display sensitivity: 0.25, 0.5, 1.0, 2.0, 3.0, 4.0 cm/mV ± 10%
  • Frequency response to screen: ESU filter out: 0.5 to 40 Hz, -3db max at 1 mVpp input; ESU filter in: 0.5 to 20 Hz -3db max at 1 mVpp input

Heart Rate Meter

  • Range: 30-300 bpm adult/child. 30-350 bpm neonatal
  • Accuracy: 30-250 bpm ± 3bpm or 3%, whichever is greater.
  • Pacer rejection: rejects all pulse of amplitude ± 2.0 mV to ± 700 mV and duration 0.1 ms to 2 ms with no tail. Rejects all pulses ± 2.0 mV to ± 700 mV and duration 0.1 ms to 2 ms with 100 ms T.C. tail <0.8 mV or 25 ms T.C. tail <0.4 mV.
  • Display update: 2 seconds ± 0.1 sec

NIBP

  • Technique: oscillometric
  • Systolic range: adult/ped mode: 40-250 mmHg, neonate mode: 15-150 mmHg.
  • Diastolic range: adult/ped mode: 30-200 mmHg, neonate mode: 10-150 mmHg.
  • Systolic accuracy: mean error less than 7 mmHg, standard deviation less than 8 mmHg (adult/pediatric). Mean error less than 3 mmHg, standard deviation less than 13 mmHg (neonatal).
  • Diastolic accuracy: mean error less than 7 mmHg, standard deviation less than 8 mmHg (adult/pediatric). Mean error less than 4 mmHg, standard deviation less than 7 mmHg (neonatal).
  • Heart rate meter range: 30-250 bpm.
  • Heart rate meter accuracy: ±3 bpm or ±3%, whichever is greater.

Pulse Oximetry

  • Accuracy (standard deviation): 70-100%, ±2%; 60-69%, ±4%; 0-59%, unspecified.
  • Response time: 13 seconds to 95% of final step change % SpO2 value from 60-95% at 72 bpm

Recorder (optional)

  • Speed: 50, 25, 12.5, 6.25, 3.125 mm/sec ±5%
  • Note: 6.25 and 3.125 are for respiration only
  • Chart size: 50 mm ±2mm

Battery Requirements

  • 2 sealed lead batteries
  • Voltage: 12 VDC
  • Capacity: 2.3 AH each
  • Run time: 2 hours fully charged battery
  • Recharge time: 16 hours maximum
  • Low battery warning: 11.6 ± 3.5%
  • Low battery cut off: 11.0 ± 2.8%