Drager Resuscitaire Infant Warmer

Drager Resuscitaire Infant Warmer

Brand: Drager
Web ID: 1431
Condition: Refurbished

0 item is in your cart. View cart now

Need help choosing?

Ask an Equipment Expert

(800) 462-8195

Listing Terms & Conditions

All product and company names are trademarks of their respective holders. Use of them does not imply any affiliation with or endorsement or sponsorship by them.

Details

THIS ITEM IS CURRENTLY UNAVAILABLE.  However for hospital use we do have newer refurbished options so please reach out to our team.  We also have infant warmers dedicated for educational and simulation purposes only.

------

The Dräger Resuscitaire infant warmer is ideal for a wide range of needs in Labor & Delivery and surgical settings. Its ease-of-use considerations, patient access, and process-driven conveniences enable caregivers to concentrate on their patients rather than their equipment. From routine warming to emergency respiratory intervention, Resuscitaire was meticulously developed to reliably meet a wide range of Labor & Delivery needs.

Features

  • Rapid Response: Progressive respiratory support capabilities help you stabilize patient breathing while safeguarding delicate airways.
  • Constant Comfort: Able to provide uniform thermoregulation across the baby's mattress and easily adapt to developmental care requirements.
  • Intuitive design
  • Precise gas delivery prevents respiratory complications in neonates.
  • Front-facing gauges
  • Choice of delivery devices

Specifications

General

  • Height: 72 in (183 cm)
  • Width: 26.5 in (68 cm)
  • Depth: 44 in (112 cm)
  • Weight: 220-280 lbs (100-127 kg)

Mattress

  • Height from floor: 39.4 in (100 cm)
  • Width: 21.0 in (53.3 cm)
  • Length: 26.0 in (66.0 cm)
  • Thickness: 1.0 in (2.54 cm)
  • Maximum capacity: 15 lbs (4.54 kg)
  • Maximum Power Requirements:
    • Model 120V: 50-60Hz, 750 W;
    • Model: 220/240V 220-240V, 50-60Hz, 750 W

Examination Light

  • Illumination >0.11 lumens/cm2 (100 ft. Candles)
  • Quartz bulb type 50 W (1,076 lux)

Controller Heating Alarms

  • Check patient: 15 minutes in Manual Mode
  • Baby temperature: ± 1° C from Set Point
  • High temperature: Skin Temp. 39.0°C ± 0.2°C
  • Probe: Short or open circuit / No probe
  • System fail: Indicates System Fail
  • Power fail: AC Power Interruption

Alarm Silence/Reset Intervals

  • Check patient: Resets clock for 15 minutes manual mode
  • Baby temperature: 10 minutes
  • High temperature: 2 minutes
  • Procedural silence: Presilences baby temp alarm for 5 minutes

Set Temperature/Heater Output

  • Set temp range: 34-38° C (>37.0° override)
  • Heater output: 0-100%, 10% increments
  • Skin temp display: 18-43° C
  • Accuracy: ± 0.2° C (31° C to 37° C)