EU Cookie Consent
We use cookies on this website to provide statistics that help us give you the best experience. By continuing to use the site without changing settings, you are agreeing to our use of cookies. Read Privacy Policy
0 item is in your cart. View cart now
All product and company names are trademarks of their respective holders. Use of them does not imply any affiliation with or endorsement or sponsorship by them.
The MAC 5500 HD Resting ECG System is a premier ECG system, delivering advanced disease management capabilities through one of the industry-leading collections of algorithms and advanced networking.
The MAC 5500 HD represents a new definition of ECG speed and accuracy, and offers:
Display
Display Type - 10.4 in (264 mm) diagonal graphics backlit color AM LCD
Display Resolution - 640 x 480 pixels with waveform enhancement
Display Data - Heart rate, patient name, ID, clock, waveforms, lead labels, speed, gain and filter settings, warning messages, prompts, and help messages
Writer
Writer Technology - Thermal dot array
Writer Speeds - 5, 12.5, 25, and 50 mm/s
Number of Traces - 3, 6, 12, or 15 user selectable
Writer Sensitivity/Gain - 2.5, 5, 10, 20, 10/5 (split calibration) mm/mV
Writer Speed Accuracy - ±2% Writer
Amplitude Accuracy - ±5%
Writer Resolution - Horizontal 1000 dpi @ 25 mm/s, 200 dpi vertical
Paper Type - Thermal, Z-fold, perforated, fan fold, 300 sheets/pack
Paper Size A Size - 8.5 in x 11 in, (214.6 mm x 280 mm) A4 Size: 8.27 in x 11.7 in (210 mm x 297.5 mm)
Keyboard
Type - Sealed elastomer with soft function keys, alphanumeric keys, writer controls, and TrimPad cursor controls
Electrical
Power Supply - AC or battery operation
Voltage - 100 to 240 VAC
Current - 0.5A @ 115 VAC, 0.3A @ 240 VAC, typical, 0.85A max
Frequency - 50 to 60 Hz
Battery Type - User replaceable, 18V @ 3.5 AH ±10% rechargeable NiMH
Battery Capacity - 100 single page reports, (typical) or six hours continuous display (without printing)
Battery Charge Time - Approximately 4.5 hours from total discharge (with display off)
Vectorcardiography
Report Formats - Vector loops of component vectors (P, QRS, ST-T)
Sensitivity - 20, 40, 80, or 160 mm/mV
Time Resolution - 2 ms
Hi-Res Late Potential Analysis and P-Wave Signal– Averaged ECG
Sensitivities:
Raw Data Template - 20 mm/mV
Average Beat - 20 mm/mV and 50 mm/mV
Filtered Signals and Vector Magnitude - 1 mm/µV
Analysis Sampling Rate - 1,000 samples/second/channel
Digital Sampling Rate - 16,000 samples/second/channel
High/Low Pass Filters Special filter using Fast Fourier Transform (FFT)
Physical Specifications
Height - 3.7 in (9.4 cm)† with display closed
Width - 15 in (38.1 cm)†
Depth - 13.8 in (35.1 cm)†
Weight - Approximately 6.8 kg (15 lbs)† including battery, without paper
Environmental Specifications
Temperature
Operating - 50°to 104° F (10° to 40° C)
Transport/Storage - –40°to 158° F (–40° to 70° C)
Humidity
Operating - 20% to 95% RH non-condensing
Transport/Storage - 15% to 95% RH non-condensing
Pressure
Operating - 700 to 1060 hPA
Transport/Storage - 500 to 1060 hPA
Magnetic Card Reader Specifications
Character Set ANSI/ISO ALPHA alphanumeric characters and ANSI/ISO BCD (subset of ASCII [ISO 646 IRV:1991])
Bar Code Scanner Specifications
Symbologies Code 39 (extended), PDF-417, Code 128, Data Matrix, Interleaved 2 of 5
Modular MAC Trolley Dimensions
Height - 37 in (94 cm)
Width - 19 in (47 cm)
Depth - 27 in (69 cm)
Height with Acquisition module holder - 59 in (150 cm)
Weight - 66 lbs. (30 kg)††
Options - Barcode scanner holder