EU Cookie Consent

We use cookies on this website to provide statistics that help us give you the best experience. By continuing to use the site without changing settings, you are agreeing to our use of cookies. Read Privacy Policy

Steris System 1

Steris System 1

Brand: Steris
Web ID: 1334
Condition: Refurbished

0 item is in your cart. View cart now

Need help choosing?

Ask an Equipment Expert

(800) 462-8195

Listing Terms & Conditions

All product and company names are trademarks of their respective holders. Use of them does not imply any affiliation with or endorsement or sponsorship by them.

Details

The Processor is an automated, microcomputer controlled device which maintains the process parameters necessary to ensure standardized and effective sterile processing. The entire STERIS PROCESS takes place within the environmentally sealed chamber of the Processor. Processing temperatures do not exceed the safe temperature limits that most manufacturers recommend to ensure protection of heat sensitive devices. Upon completion of each cycle, a comprehensive printout is produced to document the process and load information.

Various processing trays and containers permit the Processor to accommodate a wide variety of devices and procedure specific sets. The specialized trays and containers are designed to enable the operator to fix devices in the appropriate position for sterile processing, ensure a continuous exchange of sterilant and rinse on exposed surfaces of the devices (including internal structures and lumens), and protect certain types of devices from environmental contamination during transportation to sterile field following sterile processing.

Space and Weight Requirements

  • Width: 38" (97 cm.)
  • Depth: 24" (61 cm.)
  • Height: 32" (81 cm.)
  • Weight: 200 Ibs. (68 kg.)

Electrical Requirements

  • Voltage:
    • 120 VAC nominal, 60 Hz, 15A
    • 230 VAC nominal, 50 Hz, 7 A delayed fuse
  • Service:
    • 120 Volt, 20 amp., 60 Hz dedicated circuit terminated in 20 amp. Hospital grade
    • GFCI double receptacle
    • 230 Volt, 10 amp., 50 Hz dedicated circuit that is equipped with an FI safety switch (2-5 rnA)

Water Requirements

  • Pressure: 40 - 50 psig (2.8 - 3.5 bar)
  • Flow Rate: 4 gpm (15 Lpm) at 40 psig (2.8 bar)
  • Temperature: 110 - 118°F (43 - 48°C)
  • Recommended: 115 - 118°F (46 - 48°C)
  • Supply Line Size: 1/2" 1.0. minimum
  • Connection: 3/4" male hose connector
  • Quality: Potable/tap water
  • Drain: 1 1/4" 1.0. (minimum) non-backpressuring drain
  • Water usage: Approximately 13.3 gal. (50 L) per sterile processing cycle

Environmental Requirements

  • Room Temperature: 60 - 90°F (15.5 - 32°C)
  • Humidity: 10 - 90% relative, non-condensing