Details

The Valleylab Force 2 electrosurgical generator is a long standing work horse and provides cutting and coagulation in monopolar or bipolar outputs. The generator has audible and visual alarms, including REM™ Fault Indicator. The REM™ (Return Electrode Monitoring) is standard. 300 watts is available for output in the cut setting. A monopolar foot switch is included, a bipolar foot switch is optional as is a rolling cart. Refurbished units are available from Avante Health Solutions.

Force 2 Features

Blended Cut Modes

Provide flexibility through varying degrees of hemostasis. Choose one of three preset modes.

Simultaneous Independent Coagulation

Permits two surgeons to fulgurate from a single generator, for added convenience and efficacy. Accessories are activated only when keyed, reducing the potential for injury caused by inadvertent activation.

Low Voltage Coagulation

Ensures controlled, precise desiccation with less destruction of peripheral tissue, making the Valleylab Force 2 electrosurgical generator ideal for laparoscopic procedures.

REM™ Safety

Is guaranteed with Valleylab's patented adaptive REM° system, which continually monitors patient impedance levels. If a fault in the patient/return electrode contact is detected, the REM° system automatically deactivates the generator virtually eliminating the risk of burns under the return electrode. Valleylab REM° safety has been proven in more than 100 million surgical procedures worldwide.

Specifications

Physical Weight & Dimensions

  • Height: 8 inches (20 cm)
  • Width: 13 inches (33 cm)
  • Depth: 21 inches (53 cm)
  • Weight: 23 lbs (10.4 kg)

Output Waveforms

  • Cut: 510 kHz sinusoid
  • Blend 1: 510 kHz sinusoidal bursts at 50% duty cycle recurring at 31 kHz
  • Blend 2: 510 kHz sinusoidal bursts at 37.5% duty cycle recurring at 31 kHz
  • Blend 3: 510 kHz sinusoidal bursts at 25% duty cycle recurring at 31 kHz
  • Coag: 510 kHz damped sinusoidal bursts with a repetition frequency of 31 kHz Low Voltage 510 kHz sinusoidal bursts at 25%
  • Coag: duty cycle recurring at 31 kHz
  • Bipolar: 510 kHz sinusoid Output power changes by less than 5% or 5 watts, whichever is greater, as the line voltage varies from 85-135 volts (into a 300 ohm load).

Low Frequency Leakage (50-60 Hz) Source current, patient leads, all outputs tied together.

  • Normal polarity, intact chassis ground < 10 µA
  • Normal polarity, ground open <100 µA
  • Reverse polarity, ground open <100 µA
  • Sink current, 140V applied, all inputs <150 µA

High Frequency Leakage

Less than 150 mA rms

Input Power Requirements

Operating range is 85 to 135 AC volts. Current is less than 8 amperes in cut and less than 4 amperes in coag.

Power Readouts

Agree with actual power into rated load to within ± 15% or 5 watts, whichever is greater.

Adaptive REM™ System

  • Measurement Frequency: 140 kHz ± 20 kHz
  • Measurement Current: 3 mA maximum
  • Acceptable Resistance Ranges:
    • REM° pad Ñ 5-135 ohms
    • Non-REM° pad Ñ less than 20 ohms
  • Acceptance range is 5-135 ohms after REM° PolyHesive° II return electrode is applied. REM° trip is initial impedance plus 40%. For example, if the initial impedance is 30 ohms, the upper level trip is approximately 42 ohms.

Cooling

Convection, no fan.

Audio Volume

The mode indicator tones are adjustable to a minimum level of 45 dB at 1 meter. The alarm tones are not adjustable and are set at 65 dB minimum at 1 meter. Also available in a 220-240 volt, 50-60 Hz configuration. Designed to meet UL and CSA specifications.

Accessories & Supplies

Bipolar Foot Pedal

P99002

Web ID: 12630

Mobile ESU Cart

P01594

Web ID: 12631

Conmed Reusable Plate Cord-Single Plug

P00088

Web ID: 2032

Coated 3-1/2" McPherson Straight Micro (.5mm)

P01967

Web ID: 12378

Smooth 8" Bayonet Medium (2.0mm)

P00891

Web ID: 12379

Smooth 7-1/4" Bayonet Fine (1.0mm)

P01071

Web ID: 12380

Bovie Smooth 7" Straight Fine (1.0mm)

P00963

Web ID: 12381

Smooth 4-3/4" Straight Medium (2.0mm)

P00890

Web ID: 12382

Smooth 4-1/2" Curved Micro (.5mm)

P01051

Web ID: 12383

Reusable Bipolar Forcep Cord

P02039

Web ID: 12636